E2R WEIGHING RECORDS Software
Manual number:
ITKU-56-01-01-11-A
MANUFACTURER OF ELECTRONIC
WEIGHING INSTRUMENTS
RADWAG Wagi Elektroniczne, 26–600 Radom 28 Bracka Street - POLAND
Tel. +48 48 38 48 800, tel/fax. +48 48 385 00 10
Sales Department +48 48 366 80 06
www.radwag.com
Vista de pagina 0
1 2 3 4 5 6 ... 58 59

Indice de contenidos

Pagina 1 - User Manual

User Manual E2R WEIGHING RECORDS Software Manual number: ITKU-56-01-01-11-A MANUFACTURER OF ELECTRONIC WEIGHING INSTRUME

Pagina 2 - TABLE OF CONTENTS

In order to start the installation process press button „Install”. Completing installation window: After the installation is completed the compu

Pagina 3

3.1. RADWAG Synchronizer – Windows Service In the main window of E2R SYNCHRONIZER (SynchAdmin.exe), a button for changing the language and configur

Pagina 4 - 1. GENERAL INFORMATION

Stop Server Stopping the server responsible for communication with scales. Language Language selection window. Accessible languages Polish, English,

Pagina 5

3.2. Adding new scales to the Synchronizer After connecting to the database adding new scales to the Synchronizer on overlap „Scales” is possible.

Pagina 6

Scale settings: Name Any interface name Type Field to display interface type (serial, Ethernet) Config Field to display interface configuration Edi

Pagina 7 - 2. SOFTWARE INSTRUCTION

Interface settings: MediumType Connection type (None – no connection, Serial – scale connected via RS232, Ethernet – scale connected via Ethernet) C

Pagina 8

3.3. Deleting and editing of existing scale If the scale no longer works in the system it can be removed from the Synchronizer in order to avoid se

Pagina 9

17 NOTICE: Some parameters are only for service purposes and is described in the service manual. Below you can find settings of functions that are

Pagina 10 - 3. E2R SYNCHRONIZER

Synch_Towar MasaEnabled 0 Setting parameter to „1” causes enabling the update of mass standard in the scale Synch_Towar MinMaxEnabled 0 Setting param

Pagina 11

After saving go to overlap „Windows Sevice” and check if „RADWAG Synchronizer Server” operates. If the server runs you can see a visualization of

Pagina 12

2 TABLE OF CONTENTS 1. GENERAL INFORMATION ...4 1.1. Intended use ...

Pagina 13

Main window of E2R Manager: Notice: In case of lack of communication between „E2R Manager” and service Radwag Synchronizer or when the service is

Pagina 14

4.2. Show / hide the panel of connection with server In the menu Program settings of „E2R Manager” users can access the panel of connection sith se

Pagina 15

5.2. Main program window After an operator has logged in the main window appears: There are 4 areas in the main window: • Top Bar – System, •

Pagina 16

The bottom bar displayes information about the program version and logged in user their authorization. In the workspace, there are displayed al

Pagina 17

5.3.1. Database In order to display database settings, press button in module <Settings>. Functions of database settings: • Update Dat

Pagina 18

5.3.2. Updating the database After pressing button the database version will be updated. Notice: 1. Updating the database overwrites the existi

Pagina 19 - 4. E2R MANAGER

Then the window with information given earlier is displayed: After pressing OK in the window above a message box is displayed with information abo

Pagina 20

Buttons’ description: Add Adding the new group of users Delete Erasing a group of users Save Recording changes In overlap <Access Rights> all

Pagina 21 - 5. E2R WEIGHING RECORDS

After choosing a group from menu <Group> add or take off authorization according to the predetermined pattern. 5.3.6. Service settings

Pagina 22

5.4. Other settings 5.4.1. Changing the active window Only one window is to be displayed after choosing it from the list (setting a tick next to i

Pagina 23

3 6. DATABASES...37 6.1. Assortment...

Pagina 24

Accessible colour schemes: Caramel Brown Money Twins Light-blue Lilian Light-blue The Asphalt World Light-gray iMaginary Dark gray Black Black Blu

Pagina 25

5.5. Working window of the program 5.5.1. Sorting information In the working window of all accessible modules all information displayed can be fre

Pagina 26 - the connection with the

Fields: Sort Ascending Sorting from A to Z Sort Descending Sorting from Z to A Clear Sorting Disable sorting filter Group By This Column Group data

Pagina 27

5.5.3. Filter creator After choosing button „Filter editor…” you start the window of filter creator where you can freely filter data from the databa

Pagina 28

Definitions: And The condition that endorces fulfilment of all filtering criteria Or The condition that endorces fulfilment of any filtering criteri

Pagina 29

Designations: Equals Displaying only records from the selected column which are ecactly the same as the entered value Does not equal Displaying al

Pagina 30

Begins with Displaying only records from the selected column which start form the given string of characters Ends with Displaying only records from t

Pagina 31

6. DATABASES E2R Weighing Records includes following databases: • Assortment, • Contractors, • Stocks, • Packagings, • Operators. All databas

Pagina 32

6.1.1. Adding / editing assortments After pressing button a new assortment can be added to the database. After pressing button an existing assor

Pagina 33

After editing all required data save settings by pressing button <Save> and then <Close>. 6.1.2. Deleting product After pressing butt

Pagina 34

1. GENERAL INFORMATION 1.1. Intended use E2R System is a modular system for complex management of production processes that are associated in som

Pagina 35

Notice: In the working window of <Assortment> displayed information can be sorted and use additional options of filtering. See details in sect

Pagina 36

In this window you can add, erase or modify contractors that are attributed to products to be weighed. 6.2.1. Adding / editing contractors After

Pagina 37 - 6. DATABASES

6.2.2. Deleting contractors After pressing button you can delete the selected contractor fthe database. Records are protected from accidental eras

Pagina 38

6.2.4. Reports from Contractors In order to print a report from the list of contractors press button in the main window. Then a window called <

Pagina 39

6.3.1. Adding / changing stocks After pressing button you can add a new stock to the database. After pressing button you can change any data in t

Pagina 40

If the database holds weighings that are attributed to the selected stock it will not be deleted. 6.3.3. Filtering and searching stocks Eac

Pagina 41

Window of packagings: In this window you can add, erase or modify warehouses that have been added to the database. 6.4.1. Adding / changing pa

Pagina 42

Fields: Code Package code for scales Name Package name Mass Nominal package weight – tare Barcode Package Barcode All changes in the package’s data

Pagina 43

6.5.1. Adding / changing operators After pressing button you can add a new operator to the database. After pressing button you can change any dat

Pagina 44

6.5.2. Deleting operators After pressing button you can delete the selected operator from the database. Records are protected from accidental eras

Pagina 45

The network of scales can be combined with one or more types of scales. E2R SYSTEM based on Ethernet The program allows for full supervision and

Pagina 46

It allows to display all weighings collected in the SQL database. Generation of simple reports from weighings is also possible. In the main windo

Pagina 47

The tool can be started by choosing it from the pop-up menu called by right-clicking on the bar with column names. Two additional windows are displa

Pagina 48

7.2. Raport for the Materials Overlap „Material Report” allows for sorting weighings according to the product name. Different sorting criteria can

Pagina 49

Grouping sorting conditions depends on the sequence of adding them. The time interval for the report is set in the upper part of the window. Aft

Pagina 50

Grouping sorting conditions depends on the sequence of adding them. The time interval for the report is set in the upper part of the window. Aft

Pagina 51

Grouping sorting conditions depends on the sequence of adding them. The time interval for the report is set in the upper part of the window. Aft

Pagina 52

Grouping sorting conditions depends on the sequence of adding them. The time interval for the report is set in the upper part of the window. Af

Pagina 53

7.8. Archive Weighings Overlap „Archive Weighings” allows for archiving selected weigings. Accessible functions: • Move selected to Archive

Pagina 54

Restoring the archival weighings is performed similarly using button „Restore selected from the Archives”. Deleting weighings is possible ONLY aft

Pagina 55

MANUFACTURER OF ELECTRONIC WEIGHING INSTRUMENTS RADWAG WAGI ELEKTRONICZNE 26 – 600 Radom, 28 Bracka Street POLAND Tel. +48 48 38 48

Pagina 56

• processor 2 GHz or faster, • minimum 1 GB RAM (recomended 2 GB), • minimum 5 GB of free space on HD, • monitor and graphic card allowing for re

Pagina 57

Before starting cooperation with the program, each scale requires setting adequate user parameters for assuring the expected measurement acc

Pagina 58 - 8. TOOLS

In further steps follow the instructions displayed in subsequent windows. Welcome window: 8 In order to continue press „Next”. Window with ins

Pagina 59 - MANUFACTURER

Choose a folder for program (it is advisable to keep the default path),then press „Next”. Window for selecting components to be installed: 9 Cho