- 81 -
Assign <Checkweighing thresholds> option to one of the quick access
keys displayed in the bottom bar
Return to drying mode
Press < Checkweighing thresholds> soft key and insert the value of
Low Threshold and High Threshold. Return to drying mode.
After displaying <PREPARE SAMPLE> command, the display, below the
measurement result, indicates a bar graph, which using colours shows
current mass on the drying pan.
Yellow colour current mass below set Low Threshold value
Green colour current mass within set thresholds values
Red colour current mass above set value of High Threshold
2. Defining the value of target mass is carried out using a quick access key
< Target mass >
Procedure
Press grey coloured workspace, and then <Keys> option
Assign <Target value> option to one of the quick access keys displayed
in the bottom bar
Return to drying mode
Press < Target value> soft key and insert mass to be obtained in
measuring process,
Then, insert +/- percent tolerance of the target mass, (accessible range
0 100%)
The display, below the measurement result, indicates a bar graph
which contains:
o Current mass (visualized with colours)
o Target Mass value (visualized with black marker)