31
10.3. Drying chamber adjustment
Special adjustment set, which is an atypical accessory of a moisture analyzer, is
used for temperature adjustment.
1. Thermometer
2. Thermometer holder with a holder shield
Adjustment is a process that aims at calibration of temperature sensor of a
drying chamber. To proceed to temperature adjustment please insert the
temperature adjustment set into the moisture analyzer. Follow instruction below:
Step 1.
Displace:
disposable drying pan
drying pan handle
drying pan bracket
drying chamber cover
Vista de pagina 30