
CAUTION:
For drying mode Quick the user must set the drying temperature and time
interval of heating in temperature 30% higher than the specified temperature.
Time interval of heating starts at the moment when the temperature has been
reached.
For drying mode Step the user must set threshold temperature values „1” and
„2”, drying temperature and time intervals of heating for particular intermediate
temperatures (1 and 2). Time interval of heating for a given step starts at the
moment when the temperature has been reached for the given step.
14.3. Type of displayed measurement result
In order to accept carried out changes, press Print/Enter key for two times.
• Percent loss of mass
Comentarios a estos manuales