1. Thermometer
2. Thermometer holder with a shield
Adjustment is a simple process aimed at adjusting the temperature sensor
assembled in moisture analyzer’s drying chamber. Before carrying out the
temperature adjustment process, install the adjustment set inside the drying
chamber following below figures.
Step 1.
Remove from the drying
chamber the following:
Disposable pan
Pan holder
Pan bracket
Chamber cover