17.5. Error messages
Detecting of an error is signaled by the moisture analyzer by displaying a message
corresponding to the error. The message contains number of the error and a short
description informing on the possible causes of error occurrence.
List of error messages in a moisture analyzer:
Error related to data transmission.
"Range exceeded " 2.1
Maximum measuring range of a balance
exceeded.
"Range exceeded " 2.2
Maximum measuring range of a balance
exceeded.
"A/D Full" 2.4
Maximum number of divisions from
converter exceeded.
"Tare/Zero out of range " 2.5
Permissible value of zeroing or tarring
for a balance is exceeded.
"Tare out of range " 2.6
Permissible value of tarring for a balance
is exceeded.
"Result > 10% Max" 2.7
The difference between initial zero value
and the factory value is more than 10 %
from (controlled after zero initiation).
"Result > 4% Max" 2.8
The difference between the start mass
(mass of the drying pan and cross
shaped pan holder) and its value saved
in the memory is greater than 4 % on
moisture analyzer start (moisture
analyzer started with a load on the
"Difference > 1% Max" 2.9
The difference between mass of a
currently measured adjustment weight
and mass saved in balance memory is
greater than (difference >1%).
" Sample mass < 20gm " 2.10
The value of sample mass is below 20
mg.
”Sample mass out of
range "
2.11
Sample mass is below the value
determined in the parameter “Sample –
mandatory” in the database of drying
„Ref < 1000 Div” 2.12
Value of a reference mass in Percent
Setup mode in smaller than 1000
„Difference > 20%” 2.13
The adjustment coefficient of the drying
chamber differs from the factory