This module is designed according to TCP/IP 10/100 Mbit/s standard.
It comprises two signalling LED’s:
D2 lights - Ethernet connection established,
D1 blinks - transmission 10Mbit/s or 100Mbit/s.
Module accessible in two versions:
ET 1G : with external connector on the scale housing;
ET 1D : with cable (twisted pair) about 3m length led out via the gland on
the scale housing Intended for connecting to a switch (not computer).
Terminated with a standard RJ45 plug.
27.1.1. Mounting way in PUE C41H
This module is intended for mounting inside PUE C41H indicators.
It is mounted to the main board to the 10-pin J1 connector.
For ET 1G version of module a 4-pin Ethernet connector
is installed on the back wall of the housing.
For ET 1D version of module a gland is installed on the back
wall of the housing through which a shielded cable is led out
(twisted pair 3m length terminated with RJ45 plug).
Mounting procedure:
1. Unplug from mains;
2. Unscrew the back wall of the housing;
3. Install the module in J1 on the main board;
4. During installation turn your attention to plastic columns. They should
be placed one side in mounting holes in the main board and the other
side in the mounted module;
5. For ET 1G unscrew one of the plugged glands and install a socket
instead;
6. For ET 1G version connect wires from PT0017 socket to J2 and J3
connectors on the Ethernet module according to the description below;
7. For ET 1D version unplug one of the glands and led the PT0014
(twisted pair) cable out;
8. For ET 1D version connect the shield of PT0014 cable to the indicator
housing (screwed terminator, 4mm diameter);