- 35 -
14.2. ETHERNET setting
ETHERNET can be configured in <P2.4 ETHERNET>.
Inventory of default parameters:
No NAME VALUE DESCRIPTION
P2.4.1
WORK MODE SERVER, CLIENT
Ethernet connection as Server or Client.
SERVER – scale waits for connection
CLIENT – scale initiates the connection
to a HOST.
P2.4.2
IP ADDRESS 192.168.0.2 Setting an IP address.
P2.4.3
SUBNET MASK 255.255.255.0
Setting a subnet mask for Ethernet
connection.
P2.4.4
GATEWAY 192.168.0.1
Setting a gateway for Ethernet
connection.
P2.4.5
LOCAL PORT 4001
Setting a local port for Ethernet
connection. Only for devices that
work as SERVER. Servers waits
for connection on the specified port.
P2.4.6
HOST IP 192.168.0.3
Setting a host IP address (IP of a device
to connect with). Applicable only for
devices configured as CLIENTs.
P2.4.7
HOST PORT 2000
Setting a Host port (a port for connection
with a computer). Applicable only for
devices configured as CLIENTs.
P2.4.8
TIMEOUT 60
Time (in seconds) after which none-
active Ethernet connection is being
broken. Set to 0 to stop breaking the
connection.
Caution:
1. For appropriate setting of: <P2.4.2 IP ADDRESS>,
<P2.4.3 SUBNET MSK>, <P2.4.4 GATEWAY> contact
the supervisor of the net to connect with;
2. The way of connection via ETHERNET to the program
EDYTOR WPW is described in chapter 32.2.3.
3. The scale does not allow the automatic fetch of net
configuration from DHCP servers.