Radwag PUE C41H Indicator Manual de usuario Pagina 140

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 163
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 139
- 140 -
Z_A CR LF
Z_^ CR LF
- command accepted and in progress
- command comprehended but zero range overflow appeared
Z_A CR LF
Z_E CR LF
- command accepted and in progress
- time limit for stable result exceeded
Z_I CR LF
- command comprehended but cannot be executed
28.4.2. Tarring
Syntax: T CR LF
Possible answers:
T_A CR LF
T_D CR LF
- command accepted and in progress
- command completed
T_A CR LF
T_v CR LF
- command accepted and in progress
- command comprehended but tare range overflow appeared
T_A CR LF
T_E CR LF
- command accepted and in progress
- time limit for stable result exceeded
T_I CR LF
- command comprehended but cannot be executed
28.4.3. Get tare value
Syntax: OT CR LF
Possible answers:
OT_TARA CR LF
- command executed
Frame format:
1 2 3 4-12 13 14 15 16 17 18 19
O T space tare space unit space CR LF
Tare - 9 characters with decimal point justified to the right
Unit - 3 characters justified to the left