23.1. Calibration procedure
Enter <P10 USER CALIB.> according to 11.2. and then:
Unload the pan,
Press
. During adjusting a start mass you will see:
ADJ. START MASS, in the bottom line,
After completing this procedure you will see the following window:
Place the required mass on the platform 1 then press
,
During the calibration process you will see: ADJ. CALIB. FACTOR,
in the bottom line.,
After the procedure is completed you will see in: UNLOAD THE PAN
in the bottom line,
After taking off the calibration weight :
Return to weighing, saving parameters.
Return to weighing:
See 11.3.
- 89 -