Example 2:
? _ - _ _ _ _ 2 . 2 3 7 _ l b _ CR LF - the printout generated from
the scale after pressing ENTER/PRINT.
Example 3:
^ _ _ _ _ _ _ 0 . 0 0 0 _ k g _ CR LF - the printout generated from
the scale after pressing ENTER/PRINT.
28.6. Continuous transmission
The indicator can work in a continuous transmission mode. It can be
switched on or off in parameters or using RS232 commands.
Frame format sent by the indicator when continuous transmission is set:
1 2 3 4 -12 13 14 15 16 17 18
stability space sign mass space unit CR LF
Stability character
[space] if stable
[?] if not stable
[^] if an indication over the range
[v] if fan indication below the range
sign
[space] for positive values or
[-] for negative values
mass
9 characters justified to the right
unit
3 characters justified to the left
command
3 characters justified to the left
28.7. Configuring printouts
If some information included are redundant or not sufficient and there is
a necessity of changes you can design a non-standard printout. There is
possible to create up to four standard printouts (see the chapter. 17).