Radwag BALANCES 3Y SERIES Manual de usuario Pagina 120

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 246
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 119
- 120 -
Settings for header printout, weighing printout, and footer printout are
specified in point. 15.5; below description refers to setting a printout
template for report on pipette calibration.
Content of each of reports:
Working mode
User
Client
Pipette
Serial no.
No. of channels
Channel no.
Number of measurements
Operation with ISO 8655
Start date
End date
Water temperature
Temperature
Humidity
Pressure
Z coefficient
Measurements and statistics
Statistics
Status
Empty line
Dashes
Signature
Non-standard printout
24.5. Activating procedure
Carrying out pipettes calibration process requires installing an adapter for
pipettes calibration. The adapter does not come standard with the balance.
The adapter for pipettes calibration is a simple supplementary device dedicated
for calibration and checking of piston pipettes. It is designed for application on
balances MYA 21.3Y and XA 52.3Y.
The adapter has been designed to minimize evaporation occurring in liquid
handling procedures. It consists of a minimized weighing chamber installed
inside the weighing chamber of a balance. The adapter features an evaporation
ring and a dedicated weighing pan for central locating of the measuring vessel.
Application of the adapter for pipettes calibration prevents from liquid
evaporation during calibration process. Small dimensions of the adapter,
combined with the evaporation ring enable maintaining high humidity rate inside
Vista de pagina 119