Radwag BALANCES 3Y SERIES Manual de usuario Pagina 63

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 246
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 62
- 63 -
Autozero function: Dosing
The function sets autozero operation mode to default one set for
dosing mode.
Accessible values:
NO - autozero operation is automatically disabled on entering
dosing mode
YES - autozero operation is automatically enabled on entering
dosing mode
Last digit
The function determines visibility of the last decimal place
indicated on measurement result. The functions provides three
settings:
- ALWAYS: al the digits visible
- NEVER: last digit invisible
- WHEN STABLE: last digit visible only for a stable
measurement
Ambient conditions
This parameter refers to ambient conditions of the balance. By default the
parameter is set to value Stable but it is advisable to set the
parameter to value Unstable for unstable ambient conditions.
Stable setting results in faster operation, i.e. weighing takes
much less time than in case of setting to value unstable.
Unstable
Stable
Measuring units 14.2.4.
Balance user can declare in a profile: start unit, supplementary unit and two
custom measuring units.
A custom measuring unit features:
o A multiplier
o A name (3 characters)
If a custom unit is designed, then its name is added to the list of accessible
measuring units.
This menu additionally enables inserting the value of gravitational acceleration
force for balance’s place of use. It is obligatory should a balance be used to
determine mass in [N].
Vista de pagina 62
1 2 ... 58 59 60 61 62 63 64 65 66 67 68 ... 245 246

Comentarios a estos manuales

Sin comentarios