Radwag BALANCES 3Y SERIES Manual de usuario Pagina 123

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 246
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 122
- 123 -
An example of a report on pipette calibration process:
----------- Pipette calibration -----------
User Smith
Client Jones
Pipette p901\1k
Serial no. 7777
No. of channels 1
Channel no. 1
No. of measurements 10
Operation with ISO 8655 Yes
Start date 2012.03.15 07:50:44
End date 2012.03.15 07:54:34
Water temperature 22.15 °C
Temperature 21 °C
Humidity 48 %
Pressure 1005 hPa
Z coefficient 1.00328
------- Tested volume: 1000 µl -------
1 0.998 g 1000.82389 µl
2 0.998 g 1000.82389 µl
3 0.998 g 1000.82389 µl
4 0.998 g 1000.82389 µl
5 0.998 g 1000.82389 µl
6 0.998 g 1000.82389 µl
7 0.998 g 1000.82389 µl
8 0.998 g 1000.82389 µl
9 0.998 g 1000.82389 µl
10 0.998 g 1000.82389 µl
Average volume 1000.82389 µl
Average [%] 100.08 %
Systematic error 0.82389 µl
Systematic error [%] 0.08239 %
Permissible error ± 16 µl
Random error 0 µl
Random error [%] 0 %
Permissible error ± 6 µl
Status Positive
----------------------------------------
Signature
........................................
Vista de pagina 122
1 2 ... 118 119 120 121 122 123 124 125 126 127 128 ... 245 246

Comentarios a estos manuales

Sin comentarios