- 130 -
Follow above procedure and weigh the following samples.
The process can be aborted by pressing < > key.
Weighing T
On selecting the option, the balance moves to the first sample in the list,
for which weighing <T> is not completed. If there are no such samples in
a series, the balance displays a message that such process is inaccessible.
If such process is possible, then new data on initiated process is displayed
in the grey coloured workspace. During process course the bottom bar and
the workspace display data on the following steps that a user should follow.
Load a container for a sample on balance’s weighing pan and press < >
key.
On accepting by pressing < > key, the software displays a command to
unload the container from the weighing pan (displayed in the message
bar).
The user should unload the weighed container and accept the activity by
pressing < > key. The software automatically moves to the following
sample for which Weighing <T> has to be carried out.
Follow above procedure and weigh the following containers
The process can be aborted by pressing < > key.
Weighing T+ A
On selecting the option, the balance moves to the first sample in the list,
for which weighing <T> is not completed. If there are no such samples in
a series, the balance displays a message that such process is inaccessible.
If such process is possible, then new data on initiated process is displayed
in the grey coloured workspace. During process course the bottom bar and
the workspace display data on the following steps that a user should follow.
Load a container for a sample on balance’s weighing pan and press < >
key.
The mass value of a container is assigned to a sample as its tare value,
balance indication is zeroed, and a command is displayed to load a sample
in a container.
Put sample into the weighed container and accept process by pressing <
> key.
On pressing the above key, a window is displayed for inserting ambient
temperature and humidity value if option <Ambient conditions> is set to
<VALUE>, or the data on ambient conditions is automatically acquired from
the ambient conditions module and displayed for confirmation if option
<Ambient conditions> is set to <ONLINE>.