26.1. Working mode activating procedure
Procedure:
• While in the main window, press soft key located in the upper bar of the
display, which opens a submenu <Working modes> with selection of
available working modes,
• select < SQC>, mode, the software automatically returns to the main
window and displays pictogram in the upper bar,
• simultaneously, the message bar displays a command <Start control > and
a soft-key on control start is visible in the bottom tool bar:
Start control
Control start procedure:
• Starting a control process requires logging a user with corresponding access
level.
Caution:
1. Starting a control requires that an access level of the logged user is at
least a <user>. If a logged user or an anonymous user have access level
set to <guest> then on control start the software opens a message box
informing on: <Unauthorized access>.
2. Logging procedure is described in point 9 of this user manual, and
procedure determining access level for each user is provided in point 13 of
this user manual.
• Set general settings of working mode parameters to balance memory
< SQC>, description in point 26.2,
Select an appropriate product (by pressing a corresponding soft key
< > in the bottom bar or using balance’s keyboard) with correctly entered
data on a control:
Caution:
Data on a product, listed in the below table, have to be filled in before
starting a control:
Mass Nominal mass of a product
Tare Packaging mass in the adjustment unit
Min Value of –T1 error (specified in % of nominal mass)
Comentarios a estos manuales