Radwag MYA 5.3Y.F1 Microbalance Manual de usuario Pagina 47

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 198
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 46
- 47 -
Differential weighing
Means of operation: analysis of mass sample change over
time.
Statistical Quality Control
Working mode is intended to carrying out different types of
product packing processes and aimed at monitoring and / or
controlling the packing process. It enables detecting excess or
lack of product quantity in a packaging.
The settings of separate working modes include special functions specific to a
mode. They enable adapting mode’s operation to user’s individual needs. The
special settings are activated on selecting a corresponding profile. A detailed
description of special functions is provided while presenting each of working
modes.
15.1. Working mode selection
Changing working mode:
o Press name of active working mode, displayed in the left corner of the
upper bar.
o List of available working modes is displayed,
o Select name of a desired working mode, and press it.
15.2. Parameters relating to working mode
Each working mode has programmable parameters determining its
functioning. Procedure of determining these parameters:
1. Press grey coloured workspace area.
2. Below menu is displayed:
o <Settings> - additional options related to a mode,
o <Keys> - defining quick access keys,
o <Information> - selecting information displayed in the workspace,
o <Printouts> - selecting type and content of a printout,
o <Profile> - selecting a profile, which is active during balance’s
operation,
Vista de pagina 46