Radwag MYA 5.3Y.F1 Microbalance Manual de usuario Pagina 121

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 198
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 120
- 121 -
Messages displayed during a control process:
Batch quantity 1
/ 10
-
A command on process course and the quantity
of all measurements in a tested batch
Product -
Name of the controlled product
Average
-
Average mass of the controlled product
Standard
deviation
-
Standard deviation from a controlled batch
After completing the control process, the software generates a report, a the
completed control is automatically saved in the corresponding database.
Caution:
A template and an example of a report on a control process is provided in point
26.7 of this user manual.
26.4. Report on product control
An example of report on product control:
----------------- SQC ------------------
User
Product 000
Start date 2012.10.19 12:27:09
End date 2012.10.19 12:37:20
Batch number ppp
Batch quantity 10
Nominal mass 50 g
Limit T2- 5 g 10 %
Limit T1- 2.5 g 5 %
Limit T1+ 1 g 2 %
Limit T2+ 2 g 4 %
--------- Measurement 1 --------------
Net 49.998 g
--------- Measurement 2 --------------
Net 49.998 g
--------- Measurement 3 --------------
Net 50.021 g
Vista de pagina 120
1 2 ... 116 117 118 119 120 121 122 123 124 125 126 ... 197 198

Comentarios a estos manuales

Sin comentarios