Radwag MYA 5.3Y.F1 Microbalance Manual de usuario Pagina 173

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 198
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 172
- 173 -
Accessible responses:
SM_OK CR LF
SM_I CR LF
- command carried out
-
command understood but not accessible at this
moment
(e.g.: other working mode, not PARTS
COUNTING)
ES CR LF
- command not recognized (incorrect mass format)
Set target mass value (e.g. for DOSING)
Format: TV_XXXXX CR LF, where: _ - space, XXXXXmass format
Accessible responses:
TV_OK CR LF
TV_I CR LF
- command carried out
-
command understood but not accessible at this
moment (e.g.: WEIGHING mode)
ES CR LF
- command not recognized (incorrect mass format)
Set reference mass value (e.g. for PERCENT SETUP)
Format: RM_XXXXX CR LF, where: _ - space, XXXXXmass format
Accessible responses:
RM_OK CR LF
RM_I CR LF
- command carried out
-
command understood but not accessible at this
moment
(e.g.: other working mode, not PERCENT
SETUP)
ES CR LF
- command not recognized (incorrect mass format)
User log in
Format: LOGIN Name, Password CR LF (Name and password have to be
inserted as they are set in the balance small and capital letters)
Accessible responses:
LOGIN OK CR LF
- command understood, new user is logged in
LOGIN ERRROR CR LF
- command understood, an error in name or
password occurred, log in failed
ES CR LF
- command not recognized (error in format)
User log out
Format: LOGOUT CR LF
Accessible responses:
LOGOUT OK CR LF
- command understood, user is logged out
ES CR LF
- command not recognized (error in format)
Set balance mode
Format: PROFILE_Name CR LF where: _ - space (Name must be accordant
with the name form entered in a balance lowercase
and uppercase letters)
Vista de pagina 172