Accessible responses:
SM_I CR LF
command understood but not accessible at this
(e.g.: other working mode, not PARTS
- command not recognized (incorrect mass format)
Set target mass value (e.g. for DOSING)
Format: TV_XXXXX CR LF, where: _ - space, XXXXX – mass format
Accessible responses:
TV_I CR LF
command understood but not accessible at this
moment (e.g.: WEIGHING mode)
- command not recognized (incorrect mass format)
Set reference mass value (e.g. for PERCENT SETUP)
Format: RM_XXXXX CR LF, where: _ - space, XXXXX – mass format
Accessible responses:
RM_I CR LF
command understood but not accessible at this
(e.g.: other working mode, not PERCENT
- command not recognized (incorrect mass format)
User log in
Format: LOGIN Name, Password CR LF (Name and password have to be
inserted as they are set in the balance – small and capital letters)
Accessible responses:
- command understood, new user is logged in
- command understood, an error in name or
password occurred, log in failed
- command not recognized (error in format)
User log out
Format: LOGOUT CR LF
Accessible responses:
- command understood, user is logged out
- command not recognized (error in format)
Set balance mode
Format: PROFILE_Name CR LF where: _ - space (Name must be accordant
with the name form entered in a balance – lowercase
and uppercase letters)