- 167 -
SS Result release
IC Internal calibration
IC1
Lock automatic internal adjustment
IC0
Unlock automatic internal adjustment
NB
Give balance factory number
K1
Lock balance keyboard
K0
Unlock balance keyboard
BP
Start BEEP signal
OMI
List available working modes
OMS
Set working mode
OMG
Give current working mode
PC Send all implemented commands
Caution:
1. Each command must end with CR LF characters;
35.2. Format of a response to computer command
On receipt of a command, the terminal sends a response in one of the
following formats:
XX_A CR LF Command understood and in progress
XX_D CR LF
Command carried out
(appears only after the
command XX_A)
XX_I CR LF
command understood but not accessible at this
moment
XX _ ^ CR LF Command understood but max range is exceeded
XX _ v CR LF Command understood but min range is exceeded
XX _ OK CR LF Command carried out (completed)
ES_CR LF Command not recognized
XX _ E CR LF
An error occurred on command carrying out (time
limit exceeded while waiting for stable measurement
result (time limit is balance’s characteristic
parameter)
XX - stands for a name of sent command
_ - substitutes spaces
COMMANDS DESCRIPTION
Zero the balance
Format: Z CR LF