Switch off continuous transmission in current weighing unit
Format: CU0 CR LF
Accessible responses:
- command understood but not accessible at this moment
- command understood and in progress
Set low checkweighing threshold LO
Format: DH_XXXXX CR LF, where XXXXX – mass format
Accessible responses:
- command not recognized (incorrect mass format)
Set high checkweighing threshold HI
Format: UH_XXXXX CR LF, where XXXXX – mass format
Accessible responses:
- command not recognized (incorrect mass format)
Give value of low checkweighing threshold LO
Format: ODH CR LF
Accessible response: DH_MASS CR LF – command carried out
Frame format:
1 2 3 4-12 13 14 15 16 17 18 19
D H space mass space unit space CR LF
Mass - 9 characters with right justification
Unit - 3 characters with right justification
Give value of high checkweighing threshold HI
Format: OUH CR LF
Accessible response: UH_MASA CR LF – command carried out
Frame format :
1 2 3 4-12 13 14 15 16 17 18 19
U H space mass space unit space CR LF
Mass - 9 characters with right justification
Unit - 3 characters with right justification
Set mass value of a single element(only for PARTS COUNTING)
Format: SM_XXXXX CR LF, where: _ - space, XXXXX – mass format